Programmable DC High Voltage Power Supply

Programmable DC High Voltage Power Supply

The QE-HVPS-5K2M-P is a programmable DC high-voltage power supply providing a continuous 0 to +5 kV output with a maximum continuous load current of 2 mA. The unit is designed around an analog programming and monitoring interface, with dedicated voltage and current monitor outputs and hardware interlock inputs for system-level protection.

Key Applications

Key Applications

Design Intent - The architecture emphasizes stable high-voltage generation, low output ripple, analog programmability, measurement feedback, and protection against output fault events. Values below are taken from the current QUTE engineering datasheet supplied for this product.

Precision laboratory instrumentation

//01
Precision laboratory instrumentation
//01

Electron microscopes

//02
Electron microscopes
//02

Beam steering and electrostatic control

//03
Beam steering and electrostatic control
//03

Nuclear instrumentation

//04
Nuclear instrumentation
//04

Automated test equipment (ATE)

//05
Automated test equipment (ATE)
//05

Surface-science analysis equipment

//06
Surface-science analysis equipment
//06

Photomultiplier tubes (PMTs)

//07
Photomultiplier tubes (PMTs)
//07

Microchannel plates and detector systems & Mass spectrometers

//08
Microchannel plates and detector systems & Mass spectrometers
//08

Key Features

Key Features

HIGH STABILITY

<0.01%/h after warm-up
//01
HIGH STABILITY
<0.01%/h after warm-up
//01

LOW RIPPLE

<20 mV p-p at full load
//02
LOW RIPPLE
<20 mV p-p at full load
//02

ANALOG CONTROL

0 -10 V programming
//03
ANALOG CONTROL
0 -10 V programming
//03

PROTECTED

Arc / short-circuit protection
//04
PROTECTED
Arc / short-circuit protection
//04

TELEMETRY

<20 mV p-p at full load
//05
TELEMETRY
<20 mV p-p at full load
//05

COMPACT

0 -10 V programming
//06
COMPACT
0 -10 V programming
//06

HIGH STABILITY

<0.01%/h after warm-up
//01

ANALOG CONTROL

0 -10 V programming
//03

TELEMETRY

<20 mV p-p at full load
//05

LOW RIPPLE

<20 mV p-p at full load
//02

PROTECTED

Arc / short-circuit protection
//04

COMPACT

0 -10 V programming
//06

Product Specifications

Product Specifications

Specifications are the design specifications of the model QE-HVPS-5K2M-P. Always connect the earth terminal before use, and operate this high voltage equipment only with trained personnel.

Electrical Specifications


SN

Parameter

Specification

1

Primary input voltage

+24 VDC ±10%; <1 A total operational current draw

2

Rated output voltage range

0 to +5 kV DC; continuous programmable range

3

Rated output current ceiling

Maximum 2 mA total loop continuous tracking load

4

Output voltage polarity

Positive (+ve) configuration

5

Analog output programming loop

0 to +10 V control input for 0% to 100% output scaling; ±2%; Zᵢₙ = 200 kΩ

6

Voltage monitoring interface

0 to +10 V output; ±2% for 0% to 100% full-scale feedback; Zₒᵤₜ = 10 kΩ

7

Precision current monitoring

0 to +10 V output; ±2%; offset ±0.1% of FS; 0% to 100% tracking; Zₒᵤₜ = 10 kΩ

8

Input line regulation margin

<10 ppm response change per 1 V change in primary source input

9

Dynamic load regulation margin

<10 ppm response stabilization from 100 µA current draw up to maximum capacity

10

System output drift window

After 1 h warm-up: <0.01% drift/hour; <0.05% combined drift over continuous 8 h

11

Hardware output protection

Dynamic intermittent-arcing suppression circuit and short-circuit-to-ground safe trip

12

Output ripple profile (full load)

<20 mV peak-to-peak

13

Temperature coefficient

<25 ppm/°C

14

Enclosure dimensions

147 mm (W) × 98 mm (L) × 47 mm (H)

15

Total hardware unit mass

≤1.0 kg

16

Input & control connector

Integrated 10-way heavy-gauge Molex friction-lock header

17

High-voltage output

1 m double-shielded, screened coaxial high-insulation cable assembly


Functional Features


  • Continuous analog programming from 0 to +5 kV using a 0–10 V control input.

  • Dedicated voltage and current monitor outputs for closed-loop system measurement and telemetry.

  • Hardware interlock inputs for fast response to arc / short-circuit fault conditions.

  • Positive output polarity configuration as specified for QE-HVPS-5K2M-P.

  • High-stability output architecture intended for precision instrumentation.

  • Compact metal enclosure with integrated high-voltage output cable assembly.


10-Way Input & Control Interface

The unit uses an integrated 10-way Molex friction-lock connector for power, analog programming, telemetry, signal grounds, and interlock functions.


Pin

Signal

Function / Description

P1

+24VDC_IN

+24 VDC ±10% primary power input; <1 A current ceiling.

P2

PWR_GND

Primary return path / power ground for the DC input.

P3

EARTH

Low-impedance chassis earth / shield grounding connection.

P4

V_PROG_IN

0 to +10 V analog voltage programming input; Zᵢₙ = 200 kΩ.

P5

SIGD_1

Isolated instrumentation programming feedback signal ground.

P6

V_MON_OUT

0 to +10 V voltage monitor output; ±2% telemetry; Zₒᵤₜ = 10 kΩ.

P7

I_MON_OUT

0 to +10 V current monitor output; ±2%, with ±0.1% FS offset specification; Zₒᵤₜ = 10 kΩ.

P8

SIGD_2

Isolated telemetry / logging circuit reference signal ground.

P9

INTLK_A

High-speed safety arc / short fault trigger A.

P10

INTLK_B

High-speed safety arc / short fault trigger B.



High-Voltage Output


  • Output: 0 to +5 kV DC, programmable.

  • Maximum continuous output current: 2 mA.

  • HV distribution: 1 m double-shielded, screened coaxial high-insulation cable assembly.

  • Output protection includes intermittent-arcing suppression and short-circuit-to-ground safe-trip behavior.


Environmental & Mechanical


SN

Parameter

Specification

1

Operating temperature

0°C to +40°C

2

Storage temperature

−20°C to +60°C

3

Humidity

Maximum 80% RH, non-condensing

4

Cooling method

Natural convection (fanless)

5

Dimensions

147 × 98 × 47 mm (W × L × H)

6

Mass

≤1.0 kg

7

Input/control connector

10-way Molex friction-lock header

8

HV cable

1 m shielded / screened high-insulation coaxial assembly


Engineering Note

Environmental limits, cooling requirements, insulation ratings, dielectric withstand, leakage current, EMC compliance, and detailed mechanical mounting-hole geometry are not specified in the supplied QUTE engineering datasheet. They should be added once the corresponding qualification data are finalized.


Performance & System Integration Notes


Topic

Note

Programming interface

Use a low-noise 0–10 V analog source. The programming input is specified as 0–100% output scaling.

Monitoring

Voltage and current monitor outputs provide scaled 0 -10 V feedback for external ADC/DAQ/control systems.

Grounding

Use P3/EARTH for chassis/shield bonding and the specified signal-ground pins for instrumentation references.

Interlock

INTLK_A and INTLK_B are dedicated safety fault inputs associated with arc/short protection.

Warm-up

The stated output drift specification applies after a 1-hour warm-up period.

HV safety

The +5 kV output is hazardous. Enclosure access, cabling, grounding, interlocks, and discharge procedures must be handled by qualified personnel.


Ordering / Configuration


SN

Parameter

Specification

1

Product

QE-HVPS-5K2M-P

2

Output voltage

0 to +5 kV DC

3

Output current

0 to 2 mA maximum

4

Polarity

+ve

5

Control

0-10 V analog programming

6

Monitoring

0-10 V voltage and current monitor outputs

7

Input

+24 VDC

Full technical specifications, PDF

Full technical specifications, PDF


Explore Products

Explore Products

A glimpse into what’s next — bold ideas, refined execution, and forward-thinking design.